We studied the effects of newly developed aneurysm clips made of a titanium alloy (manufactured by Aesculap, A.G.) on the quality of MR and CT images. Prior to clinical application in twenty patients, the effects of five types of conventional aneurysmal clips on MR images were examined utilizing a phantom. When compared with conventional aneurysm clips, the new titanium clips produce only limited artifacts and yielded improved images in patients who underwent aneurysm clipping post-operative.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.