Objective
Performance of the Omron HBP-1320, designed for professional use by adding several features to the home blood pressure (BP) measuring device, was validated using two different protocols, specifically the ANSI/AAMI/ISO 81060-2:2013 (ANSI/AAMI/ISO) and the European Society of Hypertension International Protocol, 2010 Revision (ESH IP2), as a separate study.
Methods
Three trained medical technologists validated the performance of this device by comparing data obtained from the device with those obtained using a standard mercury sphygmomanometer throughout the study.
Results
Mean differences in mercury readings for systolic BP (SBP) and diastolic BP (DBP) between the devices were 1.6 ± 5.8 and −0.4 ± 5.3 mmHg, respectively, according to the ANSI/AAMI/ISO protocol, and mean device-observer measurement differences were −0.4 ± 4.9 and −0.2 ± 4.2 mmHg, respectively, satisfying Part 1 of ESH IP2. Differences in SBP and DBP both satisfied Part 2 of ESH IP2. The number of absolute differences in the values obtained using the device and those obtained by the observers fulfilled the requirements of the ANSI/AAMI/ISO protocol and ESH IP2.
Conclusion
Omron HBP-1320 met all requirements of the ANSI/AAMI/ISO protocol and ESH IP2.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.