A gauge block interferometer with a vacuum chamber was designed and constructed for the measurement of gauge block lengths between 10 to 1,000 mm. In the chamber, four gauge blocks under the length measurement can be mounted on the measuring table. Furthermore, the length of all four gauge blocks can be measured continuously in one measurement. In this study, four gauge blocks with the nominal length of 125, 200, 400 and 500 mm were used for the length measurement. The environment parameters in the vacuum chamber including temperature, pressure, relative humidity and CO2 concentration were measured. The refractive index of air which depends on environment parameters was determined by using updated Edlén equation. Then, the lengths of four gauge blocks were measured and found to be 125.00025, 200.00019, 400.00070 and 500.00064, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.