Abstract. Reliable, sensitive, and widely available hydrogen chloride (HCl)
measurements are important for understanding oxidation in many regions of
the troposphere. We configured a commercial HCl cavity ring-down
spectrometer (CRDS) for sampling HCl in the ambient atmosphere and developed
validation techniques to characterize the measurement uncertainties. The
CRDS makes fast, sensitive, and robust measurements of HCl in a high-finesse optical cavity coupled to a laser centred at 5739 cm−1. The accuracy was determined to reside between 5 %–10 %, calculated from laboratory and ambient air intercomparisons with annular denuders. The precision and limit of detection (3σ) in the 0.5 Hz measurement were below 6 and 18 pptv, respectively, for a 30 s integration interval in zero air. The
response time of this method is primarily characterized by fitting decay
curves to a double exponential equation and is impacted by inlet
adsorption/desorption, with these surface effects increasing with relative humidity and
decreasing with decreasing HCl mixing ratios. The minimum 90 % response
time was 10 s and the equilibrated response time for the tested inlet
was 2–6 min under the most and least optimal conditions, respectively.
An intercomparison with the EPA compendium method for quantification of
acidic atmospheric gases showed good agreement, yielding a linear
relationship statistically equivalent to unity (slope of
0.97 ± 0.15).
The CRDS from this study can detect HCl at atmospherically relevant mixing
ratios, often performing comparably or better in sensitivity, selectivity,
and response time than previously reported HCl detection methods.
Abstract. Reliable, sensitive, and widely available hydrogen chloride (HCl) measurements are important for understanding oxidation in many regions of the troposphere. We configured a commercial HCl cavity ring-down spectrometer (CRDS) for sampling HCl in the ambient atmosphere and developed calibration and validation techniques to characterize the measurement uncertainties. The CRDS makes fast, sensitive, and robust measurements of HCl in a high finesse optical cavity coupled to a laser centered at 5739 cm−1. The accuracy was determined to reside between 5–10 %, calculated from laboratory calibrations and an ambient air intercomparison with annular denuders. The precision and limit of detection (3σ) in the 0.5 Hz measurement were below 6 pptv and 18 pptv, respectively for a 30 second integration interval in zero air. The response time of this method is primarily characterized by fitting decay curves to a double exponential equation and is impacted by inlet adsorption/desorption, with these surface effects increasing with RH and decreasing with decreasing HCl mixing ratios. The response time for the tested inlet was 2–6 minutes under the most and least optimal conditions, respectively. An intercomparison with the EPA compendium method for quantification of acidic atmospheric gases showed good agreement, yielding a linear relationship statistically equivalent to unity (slope of 0.97 ± 0.15). The CRDS from this study can detect HCl at atmospherically relevant mixing ratios, often performing comparable or better in sensitivity, selectivity, and response-time from previously reported HCl detection methods.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.