Teresa A Volsko MHHS RRT FAARC, Kathy Fedor RRT-NPS, Jason Amadei RRT, and Robert L Chatburn MHHS RRT-NPS FAARC BACKGROUND: Limited data are available to describe the CPAP effects that can be expected when using high flow with a traditional nasal cannula. OBJECTIVE: To describe the relationship between the pressure generated at the airway opening and flow through a nasal cannula using a simulated infant model. We hypothesized that positive pressure generated by a standard cannula at flows > 2 L/min would be minimal and clinically unimportant. METHODS: Nares were simulated with holes drilled in a plastic fixture. A nares template for CPAP prongs served as a sizing template for the holes. Small, medium, and large nares fixtures were constructed and connected to a lung simulator that simulated spontaneous breathing. Respiratory muscle pressure was simulated by setting a waveform and adjusting the amplitude to deliver a range of tidal volumes (V T ) from 3 mL to 12 mL. Lung compliance and resistance were set at 0.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.