In this study Ru, deposited thermally on an integrated platform pyrolytic graphite tube, is proposed as a permanent modifier for the determination of Cu and Cr in urine samples by electrothermal atomic absorption spectrometry. The samples were diluted 1:1 with nitric acid (1% v/v). Pyrolysis and atomization temperatures for spiked urine samples were 1,100 degrees C and 1,900 degrees C respectively for Cu, and 1,400 degrees C and 2,500 degrees C respectively for Cr. For comparison purposes, the conventional modification with Pd+Mg was also studied. The sensitivity for Ru as permanent modifier was higher for the two analytes. The characteristic masses were 7.3 and 17.7 for Cr and Cu. The detection limits (3sigma) were 0.22 and 0.32 microg/L, for Cr and Cu, respectively. Good agreement was obtained with certified urine samples for the two elements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.