A flow coulometric electroanalytical system using a copper column electrode with a copper wire inserted into a Nafion tube was developed to determine Pb(II) content based on anodic stripping voltammetry. The electrolysis efficiency of 5 μM Pb(II) was evaluated to be 100.4±4.5 % (n=5) when the length of the copper wire and flow rate of the Pb(II) solution were 50 cm and 0.1 mL min−1, respectively. The amount of electricity due to the re‐oxidation of Pb electrodeposited at the copper column electrode was proportional to the concentration of Pb(II) in the range between 0.1 to 100 μM, and the limit of detection for Pb(II) was 0.8 μM for a deposition time of 15 min. Interference from the presence of Cd(II) could be avoided and the selective determination of Pb(II) was successfully achieved by adjustment of the electrodeposition potential.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.