We have developed a new time-resolved fluorometric (TRF) microarray detection system consisting of fluorescent NH2 nanosphere, TRF microarray detector and gamma-irradiated polystyrene chip. Using the TRF microarray detector, we detected 500 particles of the fluorescent nanosphere in one channel. Cross-talk fluorescence from the adjacent channels was little observed in the TRF microarray detector (<0.0004 %). The TRF microarray detection system was further applied for serum allergen-specific immunoglobulin E (IgE) multi-analyses. As a labeled tag antibody, an anti-human IgE Fab' fragment-conjugated fluorescent nanosphere (Fab' nanosphere) was prepared as described previously. As a chip surface appropriate for allergen immobilization, the polystyrene chip surface was modified by gamma irradiation. The immunoassay reactivity using the gamma-irradiated polystyrene chip was approximately 2.5-times improved compared with that of the non-treated polystyrene chip. Non-specific adsorption of the Fab' nanosphere onto the gamma-irradiated polystyrene chip surface was very low level (<0.0009 %). In only 20 mul of serum, six allergen-specific IgEs could be simultaneously determined in one reaction well in fewer than 90 min. Good correlation curves were obtained between the microarray immunoassay and the CAP RAST fluoro-enzyme immunoassay (CAP/RAST FEIA) method (r > 0.961). Reproducibility (CVs) of the microarray immunoassay was 8.6 % to 19.0 % (n = 5).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.