To compare accuracy, precision, and reproducibility of four commonly used myocardial T1 mapping sequences: modified Look-Locker inversion recovery (MOLLI), shortened MOLLI (ShMOLLI), saturation recovery single-shot acquisition (SASHA), and saturation pulse prepared heart rate independent inversion recovery (SAPPHIRE).
Materials andMethods:This HIPAA-compliant study was approved by the institutional review board. All subjects provided written informed consent. Accuracy, precision, and reproducibility of the four T1 mapping sequences were first compared in phantom experiments. In vivo analysis was performed in seven healthy subjects (mean age 6 standard deviation, 38 years 6 19; four men, three women) who were imaged twice on two separate days. In vivo reproducibility of native T1 mapping and extracellular volume (ECV) were measured. Differences between the sequences were assessed by using Kruskal-Wallis and Wilcoxon rank sum tests (phantom data) and mixed-effect models (in vivo data).
Results:T1 mapping accuracy in phantoms was lower with ShMOLLI (62 msec) and MOLLI (44 msec) than with SASHA (13 msec; P , .05) and SAPPHIRE (12 msec; P , .05). MOLLI had similar precision to ShMOLLI (4.0 msec vs 5.6 msec; P = .07) but higher precision than SAPPHIRE (6.8 msec; P = .002) and SASHA (8.7 msec; P , .001). All sequences had similar reproducibility in phantoms (P = .1). The four sequences had similar in vivo reproducibility for native T1 mapping (~25-50 msec; P . .05) and ECV quantification (~0.01-0.02; P . .05).
Conclusion:SASHA and SAPPHIRE yield higher accuracy, lower precision, and similar reproducibility compared with MOLLI and ShMOLLI for T1 measurement. Different sequences yield different ECV values; however, all sequences have similar reproducibility for ECV quantification.q RSNA, 2014
The CNR of simulated hypervascular liver lesions can be substantially increased and the radiation dose reduced by using an 80-kVp, high tube current CT technique.
Background-Cardiac magnetic resonance (CMR) is established for assessment of left ventricular (LV) systolic function but has not been widely used to assess diastolic function. This study tested performance of a novel CMR segmentation algorithm (LV-METRIC) for automated assessment of diastolic function. Methods and Results-A total of 101 patients with normal LV systolic function underwent CMR and echocardiography (echo) within 7 days. LV-METRIC generated LV filling profiles via automated segmentation of contiguous short-axis images (204Ϯ39 images, 2:04Ϯ0:53 minutes). Diastolic function by CMR was assessed via early:atrial filling ratios, peak diastolic filling rate, time to peak filling rate, and a novel index-diastolic volume recovery (DVR), calculated as percent diastole required for recovery of 80% stroke volume. Using an echo standard, patients with versus without diastolic dysfunction had lower early:atrial filling ratios, longer time to peak filling rate, lower stroke volume-adjusted peak diastolic filling rate, and greater DVR (all PϽ0.05). Prevalence of abnormal CMR filling indices increased in relation to clinical symptoms classified by New York Heart Association functional class (Pϭ0.04) or dyspnea (Pϭ0.006). Among all parameters tested, DVR yielded optimal performance versus echo (area under the curve: 0.87Ϯ0.04, PϽ0.001). Using a 90% specificity cutoff, DVR yielded 74% sensitivity for diastolic dysfunction. In multivariate analysis, DVR (odds ratio, 1.82; 95% CI, 1.13 to 2.57; Pϭ0.02) was independently associated with echo-evidenced diastolic dysfunction after controlling for age, hypertension, and LV mass ( 2 ϭ73.4, PϽ0.001).
Conclusions-Automated
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.