We describe a case of symptomatic focal femoral osteolysis around a screw hole distal to the hydroxyapatite-coated portion of a cannulated femoral component in a revision hip replacement. No locking screw had been inserted into this, the most proximal of the three distal holes for locking screws. The presence of polyethylene wear debris in the tissue excised from the lesion suggested that it had passed through the cannulated portion of the stem and out of the proximal unfilled distal locking hole, initiating an osteolytic reaction in an otherwise well-fixed stem. This case highlights an important design characteristic of such cannulated, uncemented femoral components. We recommend that the proximal aperture of these cannulated stems be occluded at implantation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.