Purpose: Cardiac SPECT imaging is widely used for many clinical practices such as the diagnostic of coronary artery diseases (CAD) and Myocardial Perfusion Imaging (MPI). In this work, we introduced the ProSPECT system, a dedicated cardiac SPECT system with open-gantry design. Materials and Methods: In this study we assessed the performance of the system based on NEMA-NU1-2007 standards. The ProSPECT system was characterized by measurement of planar, tomographic and clinical parameters. Results: Planar measurements showed 7.6 mm and 8.9% of spatial resolution and energy resolution at 140 keV, respectively. The tomographic resolution at 250 mm radius of rotation was 12.3 mm. Also, the maximum LOR and COR error for the tomographic imaging are obtained 1.1 mm and 1.8 mm, respectively. The clinical images that obtained using the system were confirmed by nuclear medicine physician expert. Conclusion: It concluded that the ProSPECT imaging is qualified for cardiac SPECT imaging.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.