The current endoleak classification system with some important modifications is adequate. Types I and II endoleak occur after 0 to 10% and 10% to 25% of EVARs, respectively. Many (30% to 100%) type II endoleaks will seal and have no detrimental effect, which never or rarely occurs with type I endoleaks. Not all endoleaks can be visualized with any technique, and increased pressure (endotension) can be transmitted through clot. Aneurysm pulsatility after EVAR correlates poorly with endoleaks and endotension. An enlarging aneurysm after EVAR mandates surgical or interventional treatment. These and other conclusions will help to resolve controversy and aid in the management of these vexing complications and should also point the way to future research in this field.
a sutured anastomosis fixates a graft better than any stent design tested. Hooks and barbs improve the fixation of self-expandable stents. Balloon dilatation of the proximal stent after deployment might increase fixation further. Balloon expandable stents seem to provide good fixation without the use of hooks and barbs.
Intra-aneurysm sac pressure measurement is an important adjunctive for EVAR evaluation, possibly allowing early detection of failures. High pressure is associated with AAA expansion and low pressure with shrinkage. Type II endoleaks can be responsible for AAA pressurization, and successful embolization appears to result in pressure reduction.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.