Currently, protective clothing used in clinical practice is the most representative example of efforts to reduce radiation exposure to radiation workers. However, lead is classified as a substance harmful to the human body that can cause lead poisoning. Therefore, research on the development of lead-free radiation shielding bodies is being conducted. In this study, the shielding body was manufactured by changing the size, layer, and height of the nozzle, using a 90.7% pure tungsten filament, a 3D printer material, and we compared its performance with existing protection tools. Our findings revealed that the shielding rate of the mixed tungsten filament was higher than that of the existing protective tools, confirming its potency to replace lead as the most protective material in clinical practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.