SummaryMultidimensional GC (MDGC) with macrobore WCOT and PLOT columns has shown several benefits: (1) Higherflow rates, which are used, makes possible the use of common valves as the switching device; (2) very simple system configurations are possible; (3) because the columns have relatively largecapacity, the TCD can be used to measure both organic and inorganic components; (4) micropacked or open tubular columns are precolumns which provide a wide range of selectivity; (5) both precolumn and analytical columnscan beoperatedat thesame flow rate without splitting of the sample. Four rapid sample analyses: (1) Oxygenates in gasohol; (2) natural gas; (3) refinery gas; (4) hydrocarbon types in naphtha, have been developedand performed on a less complex and economical gas chromatograph.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.