SummaryWe compared the Aura-i TM , intubating laryngeal mask airway and i-gel TM as conduits for fibreoptic-guided tracheal intubation in a manikin. Thirty anaesthetists each performed two tracheal intubations through each device, a total of 180 intubations. (22-35 [14-59]) s for the Aura-i, intubating laryngeal mask airway and i-gel, respectively. Tracheal intubation through the i-gel was the quickest (p < 0.01). Resistance to railroading of the tracheal tube over the fibrescope was significantly greater through the Aura-i compared with the intubating laryngeal mask airway and the i-gel (p = 0.001). There were no failures to intubate through the intubating laryngeal mask airway or the i-gel but six intubation attempts through the Aura-i were unsuccessful, in five owing to a railroading failure and in one owing to accidental oesophageal intubation. We conclude that the Aura-i does not perform as well as the intubating laryngeal mask airway or the i-gel as an adjunct for performing fibreoptic-guided tracheal intubation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.