PurposeTo assess the efficiency of combined use of ArcCheck® detector (AC) and portal dosimetry (PDIP) for delivery quality assurance of head and neck and prostate volumetric‐modulated arc therapy.Materials and methodsMeasurement processes were studied with the Gamma index method according to three analysis protocols. The detection sensitivity to technical errors of each individual or combined measurement processes was studied by inserting collimator, dose and MLC opening error into five head and neck and five prostate initial treatment plans. A total of 220 plans were created and 660 analyses were conducted by comparing measurements to error free planned dose matrix.ResultsFor head and neck localization, collimator errors could be detected from 2° for AC and 3° for PDIP. Dose and MLC errors could be detected from 2% and 0.5 mm for AC and PDIP. Depending on the analysis protocol, the detection sensitivity of total simulated errors ranged from 54% to 88% for AC vs 40% to 74% for PDIP and 58% to 92% for the combined process. For the prostate localization, collimator errors could be detected from 4° for AC while they could not be detected by PDIP. Dose and MLC errors could be detected from 3% and 0.5 mm for AC and PDIP. The detection sensitivity of total simulated errors ranged from 30% to 56% for AC vs 16% to 38% for PDIP and 30% to 58% for combined process.ConclusionThe combined use of the two measurement processes did not statistically improve the detectability of technical errors compared to use of single process.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.