The feasibility and accuracy of long-term transabdominal fetal electrocardiogram (fECG) recordings throughout pregnancy were studied using a portable fECG monitor. Fifteen-hour recordings of fetal heart rate (FHR) were performed in 150 pregnant women at 20-40 weeks of gestation and 1-hour recordings were performed in 22 women in labour and compared with simultaneous scalp electrode recordings. When ‡60% of fECG signals was present, the recording was defined as good. Eighty-two percent (123/150) of antenatal recordings were of good quality. This percentage increased to 90.7 (136/150 recordings) when only the night part (11 p.m.-7 a.m.) was considered. Transabdominal measurement of FHR and its variability correlated well with scalp electrode recordings (r = 0.99, P < 0.01; r = 0.79, P < 0.01, respectively). We demonstrated the feasibility and accuracy of long-term transabdominal fECG monitoring.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.