Oleamide is used as a lubricant in the manufacturing and application of polypropylene (PP) medical devices. Samples of PP were prepared with 0, 1500, and 15 000 ppm oleamide content as lubricant. The samples were either left non-sterile, sterilized with ethylene oxide (ETO), γ-radiation (γ) or autoclaved (A) and stored for up to 4 weeks. To determine the oleamide bulk-to-surface distribution depending on sterilization method and storage time an extraction method and a washing technique were applied. The oleamide content was determined by gas chromatography (GC-FID) and compared with the coefficient of friction (COF). The COF dependent on the measured lubricant content at the surface. The content of lubricant on the surface depends on the type of sterilization: ETO increased the lubricant content to some extent, γ-sterilization and autoclaving reduced it. After storage, no migration of the lubricant to the surface could be detected.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.