The leakage radiation from a diagnostic tube housing is measured "when the tube is operated at its maximum continuous rated current for the maximum rated tube potential". As the potential across the tube is reduced, the maximum rated continuous current increases to maintain a constant input in heat units per second. On several occasions the question has been asked as to the decrease in leakage radiation as the potential is reduced and the current increased. A study has been done at kilovoltages from 150 to 85 kVp at the maximum rated continuous current. The results have been related to protective barrier requirements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.