To study the effect of continuous irrigation of rotating nickel-titanium instrument with several common clinical fluids on the diameter, breaking length and breaking position of nickel-titanium instrument, so as to provide some reference and theoretical basis for clinical operation and instrument improvement.
A standardized curved root canal model was established, and ProTaper Universal (PTU) F1 instrument was selected for root canal preparation. The nickel-titanium F1 instrument was flushed with distilled water, 0.9% NaCl, 0.2% chlorhexidine, 1% sodium hypochlorite and 5% sodium hypochlorite, and the diameter, length and position of the instrument before and after breakage were recorded.
Only 5% sodium hypochlorite influenced the diameter of 6 mm marker points under different irrigation conditions (P < .05). There was no statistical difference in the length of broken instruments among all the groups, and torsional deformation mainly occurred at the end of broken instruments. The broken positions of instruments in all the groups were located at the bending segment of the root canal. The breaking frequency of the 5% sodium hypochlorite group was the highest in the area 3–5.5 mm away from apical foramen, while the other 4 groups had the highest breaking frequency in the area 0 to 1.5 mm away from apical foramen.
External irrigation with different fluids did not influence the breaking length of instruments. The closer to the apical foramen was, the higher the breaking frequency of instruments was. However, only 5% sodium hypochlorite can affect the diameter of rotary nickel-titanium instruments, and may lead to early breakage of the instrument, indicating that the use of disinfectants, except 5% sodium hypochlorite, cannot reduce breakage resistance of nickel-titanium instrument compared with distilled water flushing. Furthermore, 5% hypochlorite could not be recommended for irrigation in clinical practice.
The response surface method (RSM) is applied in the parameter optimization of nonlinear stress control tubes in power cable terminations. A local approximation method, called moving least square method (MLSM), is adopted to reduce the approximation errors. The electric field intensity is calculated by the transient finite element method. A comparison of the parameters before and after optimization reveals that the RSM is constructed efficiently and accurately by using the MLSM.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.