This study examined the characteristics of the dental X-ray fluoroscopic equipment, 'DreamRay 60F', which was recently developed in Korea. The output linearity, output reproducibility, half-value layer (HVL), leakage radiation and scattered radiation were measured using an ionisation chamber. The surface dose equivalent rate and estimated dose equivalent of the operator were also calculated. The output linearity was 0.0015-0.0175 and the coefficient of variation for the output reproducibility was 0.0013-0.0074. The experimental HVL was 2.1 mm Al, and the leakage dose rate at 100 cm from the X-ray focus ranged from 2.70 to 19.66 microGy h (-1) depending on the direction. The scattered radiation doses differed significantly (1.7-16.8 times) depending on the distance and direction. If an operator is exposed for 10 min per procedure, 5 procedures a day at 5 days a week, he/she sitting at a 90 degrees direction will receive an annual dose equivalent of 13.0 mSv (at 30 cm) and 63.7 mSv (at 50 cm) in the trunk and face surface, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.