Purpose:
To introduce a simple and inexpensive method using a fundus contact lens and a reticle, to measure retinal lesions comparing it with values obtained with high-definition optical coherence tomography (HD-OCT). This study considers optic disc as the object for comparison.
Methods:
Patients underwent routine ophthalmologic examination and their horizontal optic disc diameter was measured, both with the reticle and OCT. For measurement with reticle a simple equation was deduced, x = 0.7y, where x corresponds to the actual image size and
y
to the reticle scale reading in millimeters.
Results:
An aggregate of 127 eyes of 75 patients were dilated and examined. The calculated mean diameter according to OCT was 1.639 mm (standard deviation = 0.179) and that assenting to reticle was 1.713 mm (standard deviation = 0.175), with a difference in mean being 0.089 mm.
Conclusion:
Values obtained by this new method was found to be comparable with the OCT values for retinal measurements, useful for ophthalmologists who cannot afford expensive and sophisticated machines.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.