Abstract. Atmospheric HONO mixing ratios in indoor and outdoor environments span a
range of less than a few parts per trillion by volume (pptv) up to tens of
parts per billion by volume (ppbv) in combustion plumes. Previous HONO
calibration sources have utilized proton transfer acid displacement from
nitrite salts or solutions, with output that ranges from tens to thousands
of ppbv. Instrument calibrations have thus required large dilution flows to
obtain atmospherically relevant mixing ratios. Here we present a simple
universal source to reach very low HONO calibration mixing ratios using a
nitrite-coated reaction device with the addition of humid air and/or HCl
from a permeation device. The calibration source developed in this work can
generate HONO across the atmospherically relevant range and has high purity
(> 90 %), reproducibility, and tunability. Mixing ratios at
the tens of pptv level are easily reached with reasonable dilution flows.
The calibration source can be assembled to start producing stable HONO
mixing ratios (relative standard error, RSE ≤ 2 %) within 2 h, with output
concentrations varying ≤ 25 % following simulated transport or
complete disassembly of the instrument and with ≤ 10 % under ideal
conditions. The simplicity of this source makes it highly versatile for
field and lab experiments. The platform facilitates a new level of accuracy
in established instrumentation, as well as intercomparison studies to
identify systematic HONO measurement bias and interferences.
Abstract. Atmospheric HONO mixing ratios in indoor and outdoor environments span a range of less than a few parts per trillion by volume (pptv) up to tens of parts per billion by volume (ppbv) in combustion plumes. Previous HONO calibration sources have utilized proton transfer acid displacement from nitrite salts or solutions, with output that ranges from tens to thousands of ppbv. Instrument calibrations have thus required large dilution flows to obtain atmospherically relevant mixing ratios. Here we present a simple universal source to reach very low HONO calibration mixing ratios using a nitrite-coated reaction device with the addition of humid air and/or HCl from a permeation device. The calibration source developed in this work can generate HONO across the atmospherically relevant range and has high purity (90–99 %), reproducibility, and tunability. Mixing ratios at the tens of pptv level are easily reached with reasonable dilution flows. The calibration source can be assembled to start producing stable HONO mixing ratios (RSE ≤ 2 %) within two hours, with output concentrations varying ≤ 25 % following simulated transport or complete disassembly of the instrument, and ≤ 10 % under ideal conditions. The simplicity of this source makes it highly versatile for field and lab experiments. The platform facilitates a new level of accuracy in established instrumentation, as well as intercomparison studies to identify systematic HONO measurement bias and interferences.
Reactive nitrogen species (Nr), defined here as all N-containing compounds except N2 and N2O, have been shown to be important drivers for indoor air quality. Key Nr species include NOx...
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.