A sensing device was constructed for the amperometric determination of nitrite. It is based on the use of titanium dioxide (TiO) nanotubes template with natural fibers and carrying hemin acting as the electron mediator. A glassy carbon electrode (GCE) was modified with the hemin/TNT nanocomposite. The electrochemical response to nitrite was characterized by impedance spectroscopy and cyclic voltammetry. An amperometric study, performed at a working potential of + 0.75 V (vs. Ag/AgCl), showed the sensor to enable determination of nitrite with a linear response in the 0.6 to 130 μM concentration range and with a 59 nM limit of detection. Corresponding studies by differential study voltammetry (E = 0.75 V) exhibited a linear range from 0.6 × 10 to 7.3 × 10 M with a limit of detection of 84 nM. The sensing device was applied to the determination of nitrite in spiked tap and lake water samples. Graphical abstract Natural fibers templated synthesis of TNT immobilized hemin as electron transfer mediator for quantitative detection of nitrite with detection limit of 59 nM and good electrochemical sensitivity and the method can be used for quantitative determination of nitrite in water samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.