Reducing radiation exposure during cardiovascular catheterization is of paramount importance for both patient and staff safety. Over the years, advances in equipment and application of radiation safety protocols have significantly reduced patient dose and operator exposure. This review examines the current status of radiation protection in the cardiac and vascular catheterization laboratory and summarizes best practices for minimizing radiation exposure.
Background—
The Radiation Reduction During Cardiac Catheterization Using Real-Time Monitoring study sought to examine the effect of a radiation detection device that provides real-time operator dose reporting through auditory feedback (Bleeper Sv; Vertec Scientific Ltd; Berkshire, UK) on patient dose and operator exposure during cardiac catheterization.
Methods and Results—
Between January 2012 and May 2014, 505 patients undergoing coronary angiography, percutaneous coronary intervention, or both were randomized to use (n=253) or no use (n=252) of the Bleeper Sv radiation monitor. Operator radiation exposure was measured in both groups using a second, silent radiation exposure monitoring device. Mean patient age was 65±8 years, most patients (99%) were men, and 30% had prior coronary artery bypass graft surgery. Baseline clinical characteristics were similar in the 2 study groups. Radial access was used in 18% and chronic total occlusion percutaneous coronary intervention constituted 7% of the total procedures. Median procedure time was 17 (12–27) minutes for diagnostic angiography, 42 (28–70) minutes for percutaneous coronary intervention, and 27 (14–51) minutes in the overall study population, with similar distribution between the study groups. First (9 [4–17] versus 14 [7–25] μSv;
P
<0.001) and second (5 [2–10] versus 7 [4–14] μSv;
P
<0.001) operator radiation exposure was significantly lower in the Bleeper Sv group. Use of the device did not result in a significant reduction in patient radiation dose. The effect of the Bleeper Sv device on operator radiation exposure was consistent among various study subgroups.
Conclusions—
Use of a real-time radiation monitoring device that provides auditory feedback can significantly reduce operator radiation exposure during cardiac catheterization.
Clinical Trial Registration—
URL:
http://www.clinicaltrials.gov
. Unique identifier: NCT01510353.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.