An in-silico assessment of the performance of 3D video-fluoroscopy for the analysis of the kinematics of long bones is proposed. A reliable knowledge of in-vivo joints kinematics in physiological conditions is fundamental in the clinical field. 3D videofluoroscopy theoretically permits a mm/deg level of accuracy in joint motion analysis, but the optimization algorithm for the pose estimation is highly dependent on the geometry of the bone segment analyzed. An automated technique based on distance maps and tangency condition was applied to the elbow bones. The convergence domain was explored to quantify and optimize measurement accuracy in terms of bias and precision. By conditioning the optimization algorithm using simple image features, the estimation error had small dispersion (interquartile range within 0.5 and 0.025 mm/deg for out-of-plane and in-plane pose parameters, resp.), but with occasional bias and outliers. 3D video-fluoroscopy produced promising results for the elbow joint, but further in-vitro validation studies should be carried out.
Model-based mono-planar and bi-planar 3D fluoroscopy methods can quantify intact joints kinematics with performance/cost trade-off. The aim of this study was to compare the performances of mono- and bi-planar setups to a marker-based gold-standard, during dynamic phantom knee acquisitions. Absolute pose errors for in-plane parameters were lower than 0.6 mm or 0.6° for both mono- and bi-planar setups. Mono-planar setups resulted critical in quantifying the out-of-plane translation (error < 6.5 mm), and bi-planar in quantifying the rotation along bone longitudinal axis (error < 1.3°). These errors propagated to joint angles and translations differently depending on the alignment of the anatomical axes and the fluoroscopic reference frames. Internal-external rotation was the least accurate angle both with mono- (error < 4.4°) and bi-planar (error < 1.7°) setups, due to bone longitudinal symmetries. Results highlighted that accuracy for mono-planar in-plane pose parameters is comparable to bi-planar, but with halved computational costs, halved segmentation time and halved ionizing radiation dose. Bi-planar analysis better compensated for the out-of-plane uncertainty that is differently propagated to relative kinematics depending on the setup. To take its full benefits, the motion task to be investigated should be designed to maintain the joint inside the visible volume introducing constraints with respect to mono-planar analysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.