Immunoassay procedures have a wide application in clinical medicine and as such are used throughout clinical biochemistry laboratories both for urgent and routine testing. Clinicians and laboratory personnel are often presented with immunoassay results which are inconsistent with clinical findings. Without a high index of suspicion interferences will often not be suspected. Artifactual results can be due to a range of interferences in immunoassays which can include cross reacting substances, heterophile antibodies, autoantibodies and the high dose hook effect. Further, pre-analytical aspects and certain disease states can influence the potential for interference in immunoassays. Practical solutions for investigation of artifactual results in the setting of the routine clinical laboratory are provided.
We verified that antibody-binding substances in serum that interfere in two-site immunoassays involving murine antibodies are heterophilic antibodies. Incubation of serum containing heterophilic antibodies and a murine monoclonal antibody to human choriogonadotropin (hCG) leads to formation of a series of soluble immune complexes. We investigated the recognition of hCG by reagent antibody in the presence of heterophilic antibodies and found this recognition to be diminished. Consequently, about 30% of serum samples containing heterophilic antibodies falsely appear to contain increased concentrations of hCG. The effect on analyte recognition probably results from steric inhibition of hCG binding to complexed antibody. Heterophilic antibodies detected with a murine antibody also bound immunoglobulin from several other species but did not bind all of those tested.
Using a modified "two-site" immunoradiometric assay, termed an "interference assay," we have demonstrated the occurrence of non-analyte antibody-binding substances in approximately 40% of serum samples. These substances multivalently bind antibodies from any of several species at a site other than the antigen-binding site. Using a two-site immunoradiometric assay for human choriogonadotropin, we have investigated their effect on analyte quantification. In this system these antibody-binding substances mimic the presence of analyte; when analyte is actually present, they can also cause over- or underestimates. Addition of excess nonspecific immunoglobulin from an appropriate species eliminates this interference. However, the amount of nonspecific immunoglobulin added to an assay system may not always be enough to block interference in all samples.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.