Background/aims
To determine to what extent subjects implanted with the Argus II retinal prosthesis can improve performance compared with residual native vision in a spatial-motor task.
Methods
High-contrast square stimuli (5.85 cm sides) were displayed in random locations on a 19″ (48.3 cm) touch screen monitor located 12″ (30.5 cm) in front of the subject. Subjects were instructed to locate and touch the square centre with the system on and then off (40 trials each). The coordinates of the square centre and location touched were recorded.
Results
Ninety-six percent (26/27) of subjects showed a significant improvement in accuracy and 93% (25/27) show a significant improvement in repeatability with the system on compared with off (p<0.05, Student t test). A group of five subjects that had both accuracy and repeatability values <250 pixels (7.4 cm) with the system off (ie, using only their residual vision) was significantly more accurate and repeatable than the remainder of the cohort (p<0.01). Of this group, four subjects showed a significant improvement in both accuracy and repeatability with the system on.
Conclusion
In a study on the largest cohort of visual prosthesis recipients to date, we found that artificial vision augments information from existing vision in a spatial-motor task.
Clinical trials registry no
NCT00407602.
Reported Argus II clinical study results are in good agreement with prior in vitro and in vivo studies, and support the development of higher-density systems that employ smaller diameter electrodes. (clinicaltrials.gov identifier: NCT00407602).
Enhanced depth imaging using spectral domain OCT can be used to identify choroidal changes in inherited retinal disease. The pattern of choroidal change correlates well with the clinical appearance. It appears that the extent and pattern of choroidal thinning is dependent on the stage of the disease in some cases, and in others the causative gene defect.
The Argus™ II 60 channel epiretinal prosthesis has been developed in order to provide partial restoration of vision to subjects blinded from outer retinal degenerative disease. To date the device has been implanted in 21 subjects as part of a feasibility study. In 6 month post-implantation door finding and line tracking orientation and mobility testing, subjects have shown improvements of 86% and 73%, respectively, for system on vs. system off. In high-contrast Square Localization tests using a touch screen monitor 87% of tested subjects performed significantly better with the system on compared with off. These preliminary results show that the Argus II system provides some functional vision to blind subjects.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.