Pressure ulcer is a result of relieving pressure from skin or underlying tissues, causing localized injuries. In this study, a prototype of an electronic monitoring layer that can be placed on top of the wound patch is designed. The electronic layer is comprised of four force-sensitive pressure sensors, and an integrated temperature and humidity sensor to monitor the activities surrounding the wound site. In the simulated wound bed experiments, the results indicated that the utilization of the Bosch BME280 I2C module, when placed on top of a gauze pad, can deliver accurate and real-time monitoring of the temperature and humidity values. Furthermore, the force-sensitive resistors (FSR) installed can be utilized to detect external pressure beyond the set allowable force applied of 32 mmHg or 700g. Therefore, the electronic layer assembled from commercially available sensors can be used to monitor temperature and humidity while being able to detect externally applied pressure in real-time. However, improvements in the size and flexibility of the electronic layer are necessary to reduce the discomfort that patients suffering from pressure ulcers will experience.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.