MR imaging with a 0.02 T resistive magnet was used to establish the correlation between the histologic grading of patellar cartilage degeneration and fat water separation images or T1-and T2-relaxation times. We examined 23 cadaveric patellae. There was a positive correlation between histologically graded cartilage degeneration and TI-relaxation time. Patellar cartilage was well differentiated from surrounding structures on chemical shift water proton images, and an evaluation of cartilage degeneration was possible. No correlation was found between cartilage damage and T2-relaxation time. Chemical shift imaging at 0.02 T is easy to perform and gives further information of cartilage disorders.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.