The thermodynamic temperature of the point of inflection of the melting transition of Re-C, Pt-C and Co-C eutectics has been determined to be 2747.84 ± 0.35 K, 2011.43 ± 0.18 K and 1597.39 ± 0.13 K, respectively, and the thermodynamic temperature of the freezing transition of Cu has been determined to be 1357.80 ± 0.08 K, where the ± symbol represents 95% coverage. These results are the best consensus estimates obtained from measurements made using various spectroradiometric primary thermometry techniques by nine different national metrology institutes. The good agreement between the institutes suggests that spectroradiometric thermometry techniques are sufficiently mature (at least in those institutes) to allow the direct realization of thermodynamic temperature above 1234 K (rather than the use of a temperature scale) and that metal-carbon eutectics can be used as high-temperature fixed points for thermodynamic temperature dissemination. The results directly support the developing mise en pratique for the definition of the kelvin to include direct measurement of thermodynamic temperature.
Research into high-temperature fixed points above 1,100 • C has made significant progress since they were first reported in 1999. In particular, it has been established that single cells are repeatable at the sub-50 mK level, and intra-cell reproducibility at the 100 mK level has been demonstrated even at temperatures as high as 2,500 • C. The fixed points have been used to compare temperature and radiometry scales over a wide temperature range, and are being developed and established as secondary references for thermocouple calibrations. However, before they can be fully accepted as primary temperature references, much work remains to be done, namely:(1) Establishment of long-term stability of the fixed-point temperature; (2) Development of robust procedures for the reliable construction of the fixed-point cells (to ensure routine intra-cell reproducibility of 100 mK); (3) Demonstration of long-term robustness of the fixed-point cells; (4) Assignment of thermodynamic temperatures to a selected set of fixed points; (5) Agreement and acceptance of these temperature values by the CCT; (6) Agreed methods on how to take full utility of these new fixed points into any future International Temperature Scale (ITS)-or the current ITS-90
Below the freezing point of silver, radiation thermometers are generally calibrated by implementing the multi-point interpolation method using blackbody measurements at three or more calibration points, rather than the ITS-90 extrapolation technique. The interpolation method eliminates the need to measure the spectral responsivity and provides greater accuracy at the longer wavelengths required below the silver point. This article identifies all the sources of uncertainty associated with the interpolation method, in particular, those related to the reference blackbody temperatures (either variable-temperature or fixed-point blackbodies) and to the measured thermometer signals at these points. Estimates are given of the 'normal' and 'best' uncertainties currently achievable. A model of the thermometer response is used to propagate all the uncertainties at the reference points and provide a total uncertainty at any temperature within the calibration range. The multi-point method has the effect of constraining the total uncertainty over this range, unlike the ITS-90 technique for which the uncertainties propagate as T 2 . This article is a joint effort of the working group on radiation thermometry of the Consultative Committee for Thermometry (CCT), summarizing the knowledge and experience of all experts in this field.
The "Mise en pratique for the definition of the kelvin" (MeP-K) was established in April 2006 to be the repository of information required to perform a "practical measurement of temperature in accordance with the International System of Units (SI)." This article describes the progress made by the MeP-K HT (High Temperature Task Group) of CCT-WG5 (radiation thermometry) in drawing together the appropriate methods for accessing thermodynamic temperature above the silver point involving direct radiometric measurements on the one hand and indirect extrapolation, interpolation, and least-squares fitting on the other. An examination of the uncertainties
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.