The optical sensitivity of BeO-based luminophors has been well-known for many years. The optical stimulation of BeO with blue light is most effective. Then the dosemeters emit luminescent light in the ultraviolet-range around 325 nm. Matched on these facts a simple optically stimulated luminescence (OSL) treatment has been developed. Intense blue light-emitting diodes are used for cw-stimulation. A Hamamatsu solar blind photomultiplier detects the OSL-light. Good separation of both spectral ranges by optical filters is very important. The dosemeter has a linear dose response between approximately 20 muGy and >10 Gy. It was suggested, that a modification of stimulation conditions would allow measurements down to 1 muGy. Fading, photon energy dependence and reproducibility of OSL-signal correspond well with requirements to clinical and personal dosemeters. In addition, basic questions of the OSL-process in BeO have been investigated. A relevant point of interest was the dependency of the OSL-signal on stimulation power.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.