Background
The high failure rates of metal on metal (MoM) large diameter head total hip arthroplasty (LDH THA) and hip resurfacing (HR) prevented their long-term comparisons with regards to clinical outcome. Such knowledge would be important as ceramic LDH bearing is now available. With long-term follow-up, we investigated the difference in 1) patient-reported outcome measures (PROMs); 2) revision and adverse events rates, and 3) metal ion levels between MoM LDH THA and HR.
Methods
Forty-eight patients were randomized for LDH THA (24) or HR (24) with the same MoM articulation. At a mean follow-up of 14 years, we compared between groups different PROMs, the number of revisions and adverse events, whole blood Cobalt (Co) and Chromium (Cr) ion levels, and radiographic signs of implant dysfunction.
Results
LDH THA (all cases: revised and well-functioning) had significantly better WOMAC (94 versus 85, p = 0.04), and more frequently reported having no limitation (p = 0.04). LDH THA revision rate was 20.8% (5/24) versus 8.3% (2/24) for HR (p = 0.4). Mean Co and Cr ion levels were higher in LDH THA compared to the HR (Co: 3.8 μg/L vs 1.7 μg/L; p = 0.04 and Cr: 1.9 μg/L vs 1.4 μg/L, p = 0.1). On radiographic analyses, 2 LDH THAs showed signs of adverse reaction to metal debris, whereas 1 loose femoral HR component was documented.
Conclusion
In the long-term, MoM LDH THA had a high trunnion related revision rate but nonetheless showed better PROMs compared to HR. Provided with a well-functioning modular junction, non-MoM LDH THA would offer an appealing option.
Trial registration
ClinicalTrials.gov (NCT04516239), August 18, 2020. Retrospectively registered.
Background:
Bone stock conservation, hip anatomy preservation, and greater stability are among the promoted advantages of hip resurfacing (HR). However, the disappointing failure of some implants nearly led to its abandonment. The aim of this study was to compare clinical scores and revision and complication rates after HR with those after total hip arthroplasty (THA).
Methods:
Two hundred and three hips were randomized to 28-mm metal-on-metal (MoM) THA (99 hips) or to HR (104 hips). Main outcome measures compared between groups were the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score, the revision rate, and the complication rates. The radiographic findings were also assessed.
Results:
After a mean follow-up of 15 years (range, 14 to 16 years), 9 (4.4%) of the 203 patients were lost to follow-up and 15 (7.4%) had died. The Kaplan-Meier survivorship, with revision for any reason as the end point, was 89.2% (95% confidence interval [CI], 82.3% to 96.1%) for HR and 94.2% (95% CI, 89.3% to 99.1%) for THA (p = 0.292). The reasons for revision included infection (3 patients), recurrent dislocation (1 patient), and adverse reaction to metal debris (ARMD) (1 patient) in the THA group and ARMD (2 patients) and femoral head loosening (7 patients) in the HR group. With aseptic revision as the end point, the Kaplan-Meier survivorship was significantly higher in the THA group (97.4% versus 89.2%; p = 0.033). No dislocation occurred in the HR group compared with 4 in the THA group (p = 0.058). Both groups achieved a similar mean WOMAC score (10.7 in the HR group and 8.8 in the THA group; p = 0.749), Forgotten Joint Score (87.1 and 85.3, respectively; p = 0.410), University of California Los Angeles (UCLA) activity score (6.3 and 6.4, respectively; p = 0.189), and overall joint perception (p = 0.251).
Conclusions:
The specific HR and MoM 28-mm THA implants used in this study showed good long-term survival and function. The overall rates of complications and revisions were similar in both groups but were of different types. As it provides better femoral bone preservation and biomechanical reconstruction, HR may continue to have a role in selected patients when performed by experienced surgeons and using validated implants.
Level of Evidence:
Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.