OBJECTIVEThe objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP).RESEARCH DESIGN AND METHODSFinger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA1c and hematocrit were also measured.RESULTSThe plasma glucose range was 2.2–9.4 mmol/L with hematocrit 33–37% and mean HbA1c 5.5% ± 0.56 (SD). All meters provided plasma equivalent results except one, which reported whole blood glucose that was adjusted to plasma equivalent values. The absolute glucose difference [meter − plasma glucose] was 0.232 ± 0.69 to 0.725 ± 0.62 mmol/L mean ± SD and bias ranged from 6.1 to 15.8%. Two meters were affected by hematocrit <36% (P < 0.05).CONCLUSIONSBlood glucose meters in current use are not optimally accurate when compared with plasma glucose measurement in DP. Recognition of this deviation is essential to prevent inappropriate treatment of DP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.