Purpose: Three cases of dislocation of a Gore-Tex scleral-sutured EnVista intraocular lens are reported. The tensile strength of the lens eyelets under two suturing methods is assessed. Pursuant surgical considerations are discussed.Methods: A chart review was performed to identify cases of scleral-sutured EnVista lens dislocations. In addition, six EnVista lenses were suspended in a balanced salt solution tank, tied either with suture over haptic or simple pass technique. Eyelet tensile strength was calculated by gradual addition of weights. Eyelet fracture position and width were measured. The tensile strength of one additional EnVista lens was assessed in open air.Patients: In a retrospective, consecutive case series, three dislocated lenses were identified out of 17 surgeries from one institution. Two dislocations occurred postoperatively, and one occurred intraoperatively.Results: The EnVista eyelet demonstrated greater tensile strength tied with the simple pass method (0.27 ± 0.017 N, n = 3) than with the suture over haptic method (0.15 ± 0.016 N, n = 3; P = 0.0015). Eyelet fracture location corresponded to tensile strength. The lens in air withstood greater tensile stress.Conclusion: Stress is placed on different regions of the eyelet with each suturing method. Simple pass may withstand greater tension and decrease risk for lens fracture, but the operating surgeon must consider multiple factors when forming an operative plan.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.