The purpose of this study was to characterize the prevalence of taper damage in modular TKA components. 198 modular components were revised after 3.9±4.2y (range: 0.0–17.5y). Modular components were evaluated for fretting corrosion using a semi-quantitative 4-point scoring system. Flexural rigidity, stem diameter, alloy coupling, patient weight, age and implantation time were assessed as predictors of fretting corrosion damage. Mild-to-severe fretting corrosion (score≥2) was observed in 94/101 of the tapers on the modular femoral components and 90/97 of the modular tibial components. Mixed alloy pairs (p=0.03), taper design (p<0.001), and component type (p=0.02) were associated with taper corrosion. The results from this study supported the hypothesis that there is taper corrosion in TKA. However the clinical implications of fretting and corrosion in TKA remain unclear.
Success of pacemakers and implantable cardioverter defibrillators may be limited by premature lead failure. Lead insulation polymers, such as polyurethane (PU) and polydimethylsiloxane (PDMS), are reported to degrade over time in vivo. PU is known to undergo oxidation, whereas PDMS undergoes surface hydrolysis. Previous studies have characterized polymer degradation in vitro, in animals or in short-term human study; however, complex effects of the biochemical and mechanical environment on the lead insulation can only be fully understood by evaluating long-term-implanted leads. Therefore, we established a retrieval program to systematically characterize the chemical and surface changes in 37 of 104 retrieved pacing and defibrillator leads, implanted for ≥5 yr. Fourier transform infrared (FTIR) spectroscopy was used for chemical analysis, and a scanning electron microscope was used for surface degradation evaluation. PDMS leads were investigated for changes in the ratio of Si-O-Si to Si-C peaks, whereas PU degradation was evaluated by changes in ether (C-O-C), carbonyl (C=O), methylene (C-H), and amino (C-N/N-H) peaks. Under SEM, PDMS showed enhanced roughness but no statistical increase in Si-O-Si bonds. PU showed uniform cracking throughout the lead body and statistical changes in each of the oxidation indicative peaks. Overall, both polymers showed surface changes in the physiological environment, but PU was the only material to show chemical changes. This work is a large-scale characterization study on long-term-implanted leads that confirmed PU oxidation but not hydrolysis of PDMS in vivo. It provides important insight for manufacturers when making design improvements and for surgeons when making decisions about lead implantation.
The objective of this study was to assess the suitability of using multidirectional pin-on-disk (POD) testing to characterize wear behavior of retrieved ultrahigh molecular weight polyethylene (UHMWPE). The POD wear behavior of 25 UHMWPE components, retrieved after 10 years in vivo, was compared with 25 that were shelf aged for 10–15 years in their original packaging. Components were gamma sterilized (25–40 kGy) in an air or reduced oxygen (inert) package. 9 mm diameter pins were fabricated from each component and evaluated against CoCr disks using a super-CTPOD with 100 stations under physiologically relevant, multidirectional loading conditions. Bovine serum (20 g/L protein concentration) was used as lubricant. Volumetric wear rates were found to vary based on the aging environment, as well as sterilization environment. Volumetric wear rates were the lowest for the pins in the gamma inert, shelf aged cohort. These results support the utility of using modern, multidirectional POD testing with a physiologic lubricant as a novel method for evaluating wear properties of retrieved UHMWPE components. The data also supported the hypothesis that wear rates of gamma-inert liners were lower than gamma-air liners for both retrieved and shelf aging conditions. However, this difference was not statistically significant for the retrieved condition.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.