Ultrasonic backscatter techniques may offer a useful approach for detecting changes in cancellous bone caused by osteoporosis and other diseases. The goal of this study was to investigate the utility of a backscatter difference technique for ultrasonic bone assessment. Measurements were performed on 22 cube-shaped specimens of human cancellous bone using four broadband transducers with center frequencies 2.25, 5, 7.5, and 10 MHz. The backscatter difference spectrum D(f) was obtained by subtracting power spectra (in dB) from two different portions of the same backscatter signal. D(f) was found to be a monotonically increasing, quasi-linear function of frequency when averaged over multiple measurement sites on multiple specimens. The frequency slope of D(f) demonstrated weak to moderate correlations with specimen density (R ¼ 0.21-0.80). The frequency averaged mean of D(f) demonstrated moderate to good correlations with density (R ¼ 0.70-0.95). These results suggest that parameters based on the frequency averaged mean of the backscatter difference spectrum may be useful for bone assessment purposes.
Ultrasonic backscatter techniques are being developed to detect changes in cancellous bone caused by osteoporosis. Many techniques are based on measurements of the apparent backscatter transfer function (ABTF), which represents the backscattered power from bone corrected for the frequency response of the measurement system. The ABTF is determined from a portion of the backscatter signal selected by an analysis gate of width τw delayed by an amount τd from the start of the signal. The goal of this study was to characterize the ABTF for a wide range of gate delays (1 μs ≤ τd ≤ 6 μs) and gate widths (1 μs ≤ τw ≤ 6 μs). Measurements were performed on 29 specimens of human cancellous bone in the frequency range 1.5 to 6.0 MHz using a broadband 5-MHz transducer. The ABTF was found to be an approximately linear function of frequency for most choices of τd and τw. Changes in τd and τw caused the frequency-averaged ABTF [quantified by apparent integrated backscatter (AIB)] and the frequency dependence of the ABTF [quantified by frequency slope of apparent backscatter (FSAB)] to change by as much as 24.6 dB and 6.7 dB/MHz, respectively. τd strongly influenced the measured values of AIB and FSAB and the correlation of AIB with bone density (-0.95 ≤ R ≤ +0.68). The correlation of FSAB with bone density was influenced less strongly by τd (-0.97 ≤ R ≤ -0.87). τw had a weaker influence than τd on the measured values of AIB and FSAB and the correlation of these parameters with bone density.
Ultrasonic backscatter techniques are being developed to diagnose osteoporosis. Tissues that lie between the transducer and the ultrasonically interrogated region of bone may produce errors in backscatter measurements. The goal of this study is to investigate the effects of intervening tissues on ultrasonic backscatter measurements of bone. Measurements were performed on 24 cube shaped specimens of human cancellous bone using a 5 MHz transducer. Measurements were repeated after adding a 1 mm thick plate of cortical bone to simulate the bone cortex and a 3 cm thick phantom to simulate soft tissue at the hip. Signals were analyzed to determine three apparent backscatter parameters (apparent integrated backscatter, frequency slope of apparent backscatter, and frequency intercept of apparent backscatter) and three backscatter difference parameters [normalized mean backscatter difference (nMBD), normalized slope of the backscatter difference, and normalized intercept of the backscatter difference]. The apparent backscatter parameters were impacted significantly by the presence of intervening tissues. In contrast, the backscatter difference parameters were not affected by intervening tissues. However, only one backscatter difference parameter, nMBD, demonstrated a strong correlation with bone mineral density. Thus, among the six parameters tested, nMBD may be the best choice for in vivo backscatter measurements of bone when intervening tissues are present.
Over 52 million Americans suffer low bone mass and at least 10 million suffer from osteoporosis. This study seeks to develop a dual gate ultrasonic technique for predicting bone quality as well as bone quantity. Ultrasonic pulses from a 5 MHz transducer were propagated into regions of porous bone in 18 bone specimens from one bovine and four human donors. The dual gate technique considered the normalized mean of the backscatter difference (nMBD), which is the power difference between two gated regions of 2 μs each placed 1 μs apart over the returned signal. This ultrasonic parameter was compared to eight X-Ray MicroCT parameters describing bone quality and quantity. Among these are the Structural Model Index (SMI) and Relative Bone Volume (BV/TV). SMI grades the structure of a specimen based on its plate and rod characteristics, making it a bone quality characteristic. SMI produced an R value of 0.982 with nMBD. BV/TV, a bone quantity indicator, finds the ratio of bone volume in the total specimen volume, and showed an R value of 0.993 with nMBD.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.