Background: Biomicroscopy, B-scan ultrasound imaging, and SD-OCT are all modalities used to characterize a posterior vitreous detachment (PVD). Our objective is to assess the precision of the diagnosis of PVD by SS-OCT. Methods: This prospective observational study examines ninety-five eyes of forty-nine patients with biomicroscopy, B-scan ultrasound, and SS-OCT for the presence or absence of a complete PVD. All SS-OCT images were reviewed by two retina specialists (RWSC, ZM). All three diagnostic methods were evaluated for agreement by Cohen's kappa statistic. Results: The inter-rater reliability between retina specialists reading the SS-OCT images was 97.9% (κ = 0.957). Agreement on PVD status between SS-OCT and biomicroscopy was 85.3% (κ = 0.711). Agreement between SS-OCT and B-scan ultrasound was 83.2% (κ = 0.667). Agreement between B-scan ultrasound and biomicroscopy was 87.4% (κ = 0.743). Conclusion: For the diagnosis of complete PVD, SS-OCT allows for high accuracy and agreement between graders.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.