The described liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the detection of acrylamide in food entails aqueous room temperature extraction, SPE cleanup, and analysis by LC-MS/MS. The method is applicable to a wide variety of foods. [(13)C(3)]acrylamide is the internal standard. The limit of quantitation is 10 ppb (microg/kg). Data were obtained in duplicate from >450 products representing >35 different food types. The variability in analyte levels in certain food types suggests that it may be possible to reduce acrylamide levels in those foods.
/npsi/ctrl?action=rtdoc&an=5764712&lang=en http://nparc.cisti-icist.nrc-cnrc.gc.ca/npsi/ctrl?action=rtdoc&an=5764712&lang=fr READ THESE TERMS AND CONDITIONS CAREFULLY BEFORE USING THIS WEBSITE.http://nparc.cisti-icist.nrc-cnrc.gc.ca/npsi/jsp/nparc_cp.jsp?lang=en Vous avez des questions? Nous pouvons vous aider. Pour communiquer directement avec un auteur, consultez la première page de la revue dans laquelle son article a été publié afin de trouver ses coordonnées. Si vous n'arrivez pas à les repérer, communiquez avec nous à PublicationsArchive-ArchivesPublications@nrc-cnrc.gc.ca.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.