The jugular venous pulse (JVP) waveform provides an insight into right heart function, and its assessment is important in patients with heart failure. However, the conventional pulse-transducer (contact) method for monitoring this waveform is not frequently used because it requires a high degree of skill. The aim of this study was to confirm the effectiveness of a prototype non-contact system that employs microwave radar (24 GHz, 7 mW; non-contact system) for JVP measurement. Experiments were conducted on eight healthy male volunteers (21.88 ± 0.99 years). JVP measurements were compared between the conventional contact method and the proposed non-contact method. Change in JVP waveform was measured in response to an angle of reclining in five steps from the supine position to 75˚ of elevation. The obtained JVP measurements were similar between the two methods. Because in the non-contact method the faint pulsation of the JVP is not suppressed by the pressure of a sensor placed on the skin, the prototype microwave radar system is particularly suitable for evaluating the JVP waveform.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.