The ALTHD method may provide faster electrolyte adjustments than DUF because its efficiency depends on both the blood and dialysate flow rates. In addition, the ALTHD method is expected to provide increased efficiency. Thus, our DUF/ALTHD circuit-switching method may be clinically useful when rapid electrolyte correction is required.
Purpose
Extracorporeal circulation circuits used in cardiopulmonary bypass surgeries are increasingly being coated with polymer materials to reduce the thrombogenicity of extracorporeal devices. However, a haemoconcentrator, which corrects haematocrit and electrolyte imbalances, is not coated with polymers. In this study, we sought to assess the filtration performance of polymer-coated haemoconcentrators in order to obtain insight into their prospects for use in clinical applications.
Methods
In vitro experiments were performed to evaluate the water pressure and flow properties of polymer-coated haemoconcentrators by comparing 3 polymer-coated haemoconcentrators with 3 non-coated haemoconcentrators. The cross-sectional surfaces of both types of haemoconcentrators were observed using a scanning electron microscope (SEM).
Results
The slopes of the regression lines for estimating the filtrated fluid flow as a function of the transmembrane pressure were 6.286 ± 0.320 for polymer-coated haemoconcentrators and 3.712 ± 0.170 for non-coated haemoconcentrators. These slopes were found to be significantly different and indicate that the filtration velocity is enhanced in polymer-coated haemoconcentrators over that in non-coated haemoconcentrators. However, the hollow fibre damage observed by SEM was not shown to contribute to higher filtration flow in the polymer-coated haemoconcentrator. Taking these results into consideration, we hypothesise that a polymer coating makes a foreign surface on a hollow fibre slippery, owing to the hydrophobicity of the polymer, thereby enhancing the velocity of the filtration.
Conclusions
The results of this preliminary investigation suggest that a polymer coating can enhance the filtration performance of a haemoconcentrator and that polymer-coated haemoconcentrators might be useful in clinical applications.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.