The feasibility of cochlear stimulation by vibrating electrodes was shown although the achieved output level at frequencies ≤1 kHz was too low for EAS applications.
Introduction Most implantable hearing aids currently available were developed to compensate the sensorineural hearing loss by driving middle ear structures (e.g., the ossicles). These devices are successfully used in round window (RW) stimulation clinically, although this was initially not the intended use. Here, a novel microactuator, specifically designed for RW stimulation, was tested in human temporal bones to determine actuator performance and applicability. Methods Stapes footplate response to RW stimulation was determined experimentally in human temporal bones and the obtained sound pressure output level was estimated. Results The actuator had a flat displacement response between 0.125 and 4 kHz, a resonance between 4 and 7 kHz, and a roll-off above. At increasing contact force, the stapes footplate displacement decreased by 5–10 dB re μm for forces ≥ 2 mN. The equivalent sound pressure level between 0.125 and 4 kHz amounted to 87–97 eq dB SPL and increased to 117 eq dB SPL for frequencies of 4–7 kHz. The total harmonic distortion (THD) of the actuator ranged within 15–40% for static forces of 5 mN. Conclusion The feasibility of an electromagnetic actuator that may be placed into the RW niche was demonstrated but requires further optimization in terms of THD and static force sensitivity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.