Diabetes patients are increasingly using a continuous glucose sensor to monitor blood glucose and an insulin pump connected to an infusion cannula to administer insulin. Applying these devices requires two separate insertion sites, one for the sensor and one for the cannula. Integrating sensor with cannula to perform glucose sensing and insulin infusion through a single insertion site would significantly simplify and improve diabetes treatment by reducing the overall system size and the number of necessary needle pricks. Presently, several research groups are pursuing the development of combined glucose sensing and insulin infusion devices, termed single-port devices, by integrating sensing and infusion technologies created from scratch. Methods: Instead of creating the device from scratch, we utilized already existing technologies and introduced three design concepts of integrating commercial glucose sensors and infusion cannulas. We prototyped and evaluated each concept according to design simplicity, ease of insertion, and sensing accuracy. Results: We found that the best single-port device is the one in which a Dexcom sensor is housed inside a Medtronic cannula so that its glucose sensitive part protrudes from the cannula tip. The low degree of component modification required to arrive at this configuration allowed us to test the efficiency and safety of the device in humans. Conclusion: Results from these studies indicate the feasibility of combining commercial glucose sensing and insulin delivery technologies to realize a functional single-port device. Significance: Our development approach may be generally useful to provide patients with innovative medical devices faster and at reduced costs.
Aim: To determine the effect of insulin on the resistance of subcutaneous tissue to the flow of infusion fluids.Materials and methods: Thirty subjects with type 1 diabetes wore two Accu-Chek Spirit Combo insulin pumps with Accu-Chek FlexLink infusion sets (Roche Diabetes Care, Mannheim, Germany) for 7 days. One pump was filled with insulin aspart (Novo Nordisk, Bagsvaerd, Denmark) and used for continuous subcutaneous insulin infusion (CSII). The other pump was filled with insulin diluting medium (IDM; Novo Nordisk) and used to deliver IDM subcutaneously at rates identical to those employed for CSII.Both infusion sites were assessed daily by measuring the pressure required to infuse various bolus amounts of IDM.Results: On day 1, maximum pressure (P max ) and tissue flow resistance (TFR; calculated from measured pressure profiles) were similar for both infusion sites (P > 0.20).During the subsequent study days, the P max and TFR values observed at the IDM infusion site remained at levels comparable to those seen on day 1 (P > 0.13). However, at the site of CSII, P max and TFR progressively increased with CSII duration. By the end of day 7, P max and TFR reached 25.8 */2.11 kPa (geometric mean */geometric standard deviation) and 8.64 */3.48 kPa*s/μL, respectively, representing a remarkable 3.5-and 20.6-fold increase relative to the respective P max and TFR values observed on day 1 (P < 0.001).
Conclusion:Our results suggest that insulin induces a progressive increase in the resistance of subcutaneous tissue to the introduction of fluid; this has important implications for the future design of insulin pumps and infusion sets.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.