In bioelectrical potential measurement with restricted skin-electrode contact area, such as in dense array EEG measurement where the electrolyte bridging effect 1 is a major concern for signal reliability, an enhanced electrolyte solution is required for the skinelectrode impedance to reach the sufficiently low level within the minimum time interval. In this study, an electrolyte gel with its skin permeation ability enhanced by ethanol or propylene glycol has been investigated. The standard skin-electrode impedance measurement was carried out on the forehead in an area of 6 mm in diameter using standard Ag/AgCl EEG electrodes. The gel solutions with 0%, 7%, 18% and 28% of enhancers by volume are compared. The results show that both ethanol and propylene glycol reduce the permeation barrier of the stratum corneum so that ions in the electrolyte gel can penetrate more readily into the skin, enhancing the skin-electrode conductivity in reaching the steady value at a faster rate. It was further found that for the gel with higher percentage of ethanol, lower minimum skin-electrode impedance value was obtained. However, as the percentage of propylene glycol increased, it fails to attain low steady impedance values in the skin-electrode impedance measurements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.