Transdermal drug delivery TDD systems have many advantages but are conventionally limited by the low permeability of skin. The idea of using microneedles to painlessly penetrate the topmost impermeable stratum corneum has previously been put forward. In this paper, the fabrication of solid and hollow silicon microneedles with straight side-walls and with the following dimensions: 20-100 microm in diameter and 100-150 microm in length is described. In vitro tests demonstrate that with prior solid microneedle application, transdermal drug transport is significantly increased by 10-20 times, with the degree of enhancement being related to needle diameter. In vivo tests in diabetic animals, however, were unable to demonstrate any delivery of insulin through the hollow microneedles. It is proposed that two factors, microneedle length and tip sharpness, have to be improved for systemic drug delivery to be seen in vivo.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.