We hypothesized that the posterior-to-anterior (PA) calcaneal interlocking screw of the Biomet Ankle Arthrodesis Nail would increase rotational stability secondary to increased bone purchase compared with the standard lateral-to-medial (transverse) screw. Each of 10 fresh human cadaver lower limbs (five matched pairs) were stabilized with a nail inserted retrograde through the calcaneus, talus, and tibia according to standard technique. One limb of each pair was fixed with a transverse calcaneal screw; the contralateral limb, with a PA calcaneal screw. Each limb was then subjected to torsional testing on an MTS Mini Bionix load frame. The PA screw construct was significantly stiffer than the transverse screw construct: 1.96 and 1.41 Nm/E, respectively (P < 0.036).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.