Purpose: To present strategies for optimization of lens power formula constants and to show options how to present the results adequately.
Methods: A dataset of N=1601 preoperative biometric values, lens power data and postoperative refraction data was split into a training set and a test set using a random sequence. Based on the training set we calculated the formula constants for established lens calculation formulae with different methods. Based on the test set we derived the formula prediction error as difference of the achieved refraction from the formula predicted refraction.
Results: For formulae with 1 constant it is possible to back-calculate the individual constant for each case using formula inversion. However, this is not possible for formulae with more than 1 constant. In these cases, more advanced concepts such as nonlinear optimization strategies are necessary to derive the formula constants. During cross-validation, measures such as the mean absolute or the root mean squared prediction error or the ratio of cases within mean absolute prediction error limits could be used as quality measures.
Conclusions: Different constant optimization concepts yield different results. To test the performance of optimized formula constants a cross-validation strategy is mandatory. We recommend performance curves, where the ratio of cases within absolute prediction error limits is plotted against the mean absolute prediction error.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.