This study aims to evaluate whether optical coherence tomography (OCT) using both the surface and the endoluminal technique is feasible to investigate the locations and degree of encrustation process in clinically used ureteral stents. After removal from patients, 14 polyurethane JJ stents were investigated. A fresh JJ served as a control. The external surfaces were examined using an endoscopic surface OCT whereas the intraluminal surfaces were investigated by an endoluminal radial OCT device. The focus was on detection of encrustation or crystalline sedimentation. In 12 female and two male patients, the median indwelling time of the ureteral catheter was 100 days (range, 19-217). Using the endoluminal OCT, the size and grade of intraluminal encrustation could be expressed as a percentage relating to the open lumen of the reference stent. The maximum encrustation observed resulted in a remaining unrestricted lumen of 15-35 % compared to the reference. The luminal reduction caused by encrustation was significantly higher at the proximal end of the ureteral stent as compared to its distal part. The extraluminal OCT investigations facilitated the characterization of extraluminal encrustation. OCT techniques were feasible and facilitated the detection of encrustation of double pigtail catheters on both the extra and intra luminal surface. Quantitative expression of the degree of intraluminal encrustation could be achieved, with the most dense and thickened occurrence of intraluminal incrustation in the upper curl of the JJ stent.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.