The objective for this work was to develop a commissioning methodology for the treatment delivery components of the AccuBoost system, as well as to establish a routine quality assurance program and appropriate guidance for clinical use based on the commissioning results. Various tests were developed: 1) assessment of the accuracy of the displayed separation value; 2) validation of the dwell positions within each applicator; 3) assessment of the accuracy and precision of the applicator localization system; 4) assessment of the combined dose profile of two opposed applicators to confirm that they are coaxial; 5) measurement of the absolute dose delivered with each applicator to confirm acceptable agreement with dose based on Monte Carlo modeling; 6) measurements of the skin‐to‐center dose ratio using optically stimulated luminescence dosimeters; and 7) assessment of the mammopad cushion's effect on the center dose. We found that the difference between the measured and the actual paddle separation is <0.1 cm for the separation range of 3 cm to 7.5 cm. Radiochromic film measurements demonstrated that the number of dwell positions inside the applicators agree with the values from the vendor, for each applicator type and size. The shift needed for a good applicator‐grid alignment was within 0.2 cm. The dry‐run test using film demonstrated that the shift of the dosimetric center is within 0.15 cm. Dose measurements in water converted to polystyrene agreed within 5.0% with the Monte Carlo data in polystyrene for the same applicator type, size, and depth. A solid water‐to‐water (phantom) factor was obtained for each applicator, and all future annual quality assurance tests will be performed in solid water using an average value of 1.07 for the solid water‐to‐water factor. The skin‐to‐center dose ratio measurements support the Monte Carlo‐based values within 5.0% agreement. For the treatment separation range of 4 cm to 8 cm, the change in center dose would be <1.0% for all applicators when using a compressed pad of 0.2 cm to 0.3 cm. The tests performed ensured that all treatment components of the AccuBoost system are functional and that a treatment plan can be delivered with acceptable accuracy. Based on the commissioning results, a quality assurance manual and guidance documents for clinical use were developed.PACS numbers: 87.55.Qr, 87.56.Da, 87.90.+y
Purpose: To implement an efficient and robust process for AccuBoost planning and treatment delivery that can be safely performed by a single Physicist while minimizing patient's total session time. Methods: Following a thorough commissioning and validation process, templates were created in the brachytherapy planning system for each AccuBoost applicator. Tables of individual and total nominal dwell times for each applicator as a function of separation were generated to streamline planning while an Excel‐based nomogram provided by the vendor functions as a secondary verification of the treatment parameters. Tables of surface dose as a function of separation and applicator, along with concise guidance documents for applicator selection, are readily available during the planning process. The entire process is described in a set of detailed Standard Operating Procedures which, in addition to the items described above, include a verbal time‐out between the primary planner and the individual performing the secondary verification as well as direct visual confirmation of applicator placement using an articulated mirror. Prior to treatment initiation, a final time‐out is conducted with the Radiation Oncologist. Chart documentation is finalized after the patient is released from compression following completion of the treatment. Results: With the aforementioned procedures, it has been possible to consistently limit the time required to prepare each treatment such that the patient is typically under compression for less than 10 minutes per orientation prior to the initiation of the treatment, which is particularly important for APBI cases. This process can be overseen by a single physicist assisted by a dosimetrist and has been optimized during the past 16 months, with 180 treatment sessions safely completed to date. Conclusion: This work demonstrates the implementation of an efficient and robust process for real‐time‐planned AccuBoost treatments that effectively minimizes the time that the patient is under compression.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.