This study aimed to compare the mechanical properties of differently-tapered EndoWave nickel-titanium endodontic rotary instruments (size #30/0.04 taper and #30/0.06 taper; Group 0.04 and 0.06, respectively). Torsional and bending properties were evaluated with the cyclic fatigue test and the cantilever bending test, respectively. Canal transportation was assessed following instrumentation of curved canals in resin blocks. Torque and apical force during instrumentation were determined using an automated instrumentation and torque/force analyzing device. The number of cycles to fracture was Group 0.04>Group 0.06 (p<0.05). The bending load values were Group 0.06>Group 0.04 (p<0.05). Group 0.04 better maintained the original canal shape compared with Group 0.06. The median clockwise torque value was Group 0.06>Group 0.04 (p<0.05), whereas no significant differences were detected in the median counterclockwise torque and apical force values (p>0.05). Under the present experimental condition, 0.04- and 0.06-tapered EndoWave instruments exhibited different mechanical properties.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.